Wall-correction and absorbed-dose conversion factors for Fricke dosimetry: Monte Carlo calculations and measurements

Chang ming Ma, A. E. Nahum, D. W.O. Rogers, K. R. Shortt, C. K. Ross, A. F. Bielajew

Research output: Contribution to journalArticlepeer-review

35 Scopus citations

Abstract

For megavoltage radiotherapy photon beams, EGS4 Monte Carlo calculations show, and experimental measurements confirm with an accuracy of 0.2%, that glass or quartz-walled vials used in Fricke dosimetry increase the dose in the Fricke solution. This is mainly caused by increased electron scattering from the glass which increases the dose to the Fricke solution. The dose perturbation is shown to vary from nothing in a 60Co beam up to 2% in a 24-MV beam. For plastic vials of similar shapes, calculations demonstrate that the effect is in the opposite direction and even at high energies it is much less (0.2% to 0.5%).

Original languageEnglish
Pages (from-to)283-292
Number of pages10
JournalMedical Physics
Volume20
Issue number2
DOIs
StatePublished - Mar 1993