Abstract
Current gated radiation therapy starts with simulation 4DCT images of a patient with lung cancer. We propose a method to confirm the phase of 4DCT for planning and setup position at the time of treatment. An intensity-based rigid algorithm was developed in this work to register an orthogonal set of on-board projection X-ray images with each phase of the 4DCT. Multiple DRRs for one of ten 4DCT phases are first generated and the correlation coefficient (CC) between the projection X-ray image and each DRR is computed. The maximum value of CC for the phase is found via a simulated annealing optimization process. The whole process repeats for all ten phases. The 4DCT phase that has the highest CC is identified as the breathing phase of the X-ray. The phase verification process is validated by a moving phantom study. Thus, the method may be used to independently confirm the correspondence between the gating phase at the times of 4DCT simulation and radiotherapy delivery. When the intended X-ray phase and actual gating phase are consistent, the registration of the DRRs and the projection images may also yield the values of patient shifts for treatment setup. This method could serve as the 4D analog of the conventional setup film as it provides both verification of the specific phase at the time of treatment and isocenter positioning shifts for treatment delivery.
Original language | English |
---|---|
Pages (from-to) | 117-125 |
Number of pages | 9 |
Journal | Physica Medica |
Volume | 26 |
Issue number | 3 |
DOIs | |
State | Published - 2010 |
Keywords
- 4DCT
- Registration
- Setup verification