TY - JOUR
T1 - Photon beam characterization and modelling for Monte Carlo treatment planning
AU - Deng, Jun
AU - Jiang, Steve B.
AU - Kapur, Ajay
AU - Li, Jinsheng
AU - Pawlicki, Todd
AU - Ma, C. M.
PY - 2000/2
Y1 - 2000/2
N2 - Photon beams of 4, 6 and 15 MV from Varian Clinac 2100C and 2300C/D accelerators were simulated using the EGS4/BEAM code system. The accelerators were modelled as a combination of component modules (CMs) consisting of a target, primary collimator, exit window, flattening filter, monitor chamber, secondary collimator, ring collimator, photon jaws and protection window. A full phase space file was secondary directly above the upper photon jaws and analysed using beam data processing software, BEAMDP, to derive the beam characteristics, such as planar fluence, angular distribution, energy spectrum and the fractional contributions of each individual CM. A multiple- source model has been further developed to reconstruct the original phase space. Separate sources were created with accurate source intensity, energy, fluence and angular distributions for the target, primary collimator and flattening filter. Good agreement (within 2%) between the Monte Carlo calculations with the source model and those with the original phase space was achieved in the dose distributions for field sizes of 4 cm x 4 cm to 40 cm x 40 cm at source surface distances (SSDs) of 80-120 cm. The dose distributions in lung and bone heterogeneous phantoms have also been found to be in good agreement (within 2%) for 4, 6 and 15 MV photon beams for various field sizes between the Monte Carlo calculations with the source model and those with the original phase space.
AB - Photon beams of 4, 6 and 15 MV from Varian Clinac 2100C and 2300C/D accelerators were simulated using the EGS4/BEAM code system. The accelerators were modelled as a combination of component modules (CMs) consisting of a target, primary collimator, exit window, flattening filter, monitor chamber, secondary collimator, ring collimator, photon jaws and protection window. A full phase space file was secondary directly above the upper photon jaws and analysed using beam data processing software, BEAMDP, to derive the beam characteristics, such as planar fluence, angular distribution, energy spectrum and the fractional contributions of each individual CM. A multiple- source model has been further developed to reconstruct the original phase space. Separate sources were created with accurate source intensity, energy, fluence and angular distributions for the target, primary collimator and flattening filter. Good agreement (within 2%) between the Monte Carlo calculations with the source model and those with the original phase space was achieved in the dose distributions for field sizes of 4 cm x 4 cm to 40 cm x 40 cm at source surface distances (SSDs) of 80-120 cm. The dose distributions in lung and bone heterogeneous phantoms have also been found to be in good agreement (within 2%) for 4, 6 and 15 MV photon beams for various field sizes between the Monte Carlo calculations with the source model and those with the original phase space.
KW - Models, Theoretical
KW - Monte Carlo Method
KW - Particle Accelerators
KW - Phantoms, Imaging
KW - Photons/therapeutic use
KW - Radiotherapy Planning, Computer-Assisted
KW - Software
UR - http://www.scopus.com/inward/record.url?scp=0033962673&partnerID=8YFLogxK
UR - https://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=purepublist2023&SrcAuth=WosAPI&KeyUT=WOS:000085410400013&DestLinkType=FullRecord&DestApp=WOS
U2 - 10.1088/0031-9155/45/2/311
DO - 10.1088/0031-9155/45/2/311
M3 - Article
C2 - 10701512
SN - 0031-9155
VL - 45
SP - 411
EP - 427
JO - Physics in Medicine and Biology
JF - Physics in Medicine and Biology
IS - 2
ER -