T1 - Initial Clinical Experience Treating Patients With Lung Cancer on a 6MV-Flattening-Filter-Free O-Ring Linear Accelerator
AU - Barsky, AR
AU - Lin, H
AU - Mendes, A
AU - Dreyfuss, A
AU - Wright, C
AU - Anstadt, EJ
AU - Berman, AT
AU - Levin, WP
AU - Cengel, KA
AU - Anderson, N
AU - Dong, L
AU - Metz, JM
AU - Li, TR
AU - Feigenberg, S
N1 - Copyright © 2020, Barsky et al.
PY - 2020/9/9
Y1 - 2020/9/9
N2 - Introduction Modern technologies, like intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT), have improved the therapeutic ratio of thoracic radiotherapy (TRT) for lung cancer (LC). Halcyon™ (Varian Medical Systems, Palo Alto, CA, USA), a novel 6MV-flattening-filter-free O-ring linear accelerator (6X-FFF ORL), was designed to deliver IMRT and VMAT with greater speed than a C-arm linac. Herein, we report our initial clinical experience treating patients with LC on this linac. Methods All patients who received TRT for LC on the 6X-FFF ORL at our institution were retrospectively identified. Patients' clinicopathologic data, radiotherapy details, early disease-control and toxicity outcomes, dosimetric data, couch corrections, and treatment times are reported. Results Between 10/2018-12/2019, 30 consecutive patients (median age 66 years, range 54-94 years) received definitive or post-operative TRT for LC (median 66 Gy/33 fractions; range 5-70 Gy/2-37 fractions) following four-dimensional computed tomography (CT) simulation (97%) using daily kilovoltage KV cone-beam CT (CBCT) (100%) on a 6X-FFF ORL for non-small cell LC (84%) or small cell LC (16%), with 53% receiving VMAT, 43% receiving static-field IMRT, and 77% receiving concurrent systemic therapy. All plans were approved through institutional peer review. The average three-dimensional vector couch correction based on CBCT guidance was 0.90 ± 0.50 cm. The average beam-on and beam on plus CBCT times were 1.7 ± 1.1 min, and 5.0 ± 3.2 min, respectively. Grade 3 dyspnea and fatigue occurred in 3% and 3% of patients, respectively. There were no grade ≥4 toxicities. Conclusion In this first clinical report of TRT for LC on a 6X-FFF ORL, daily CBCT-guided treatment was fast and safe with respect to dosimetry and clinical outcomes. Thus, use of this linac for TRT may increase LC patient throughput without a detriment in radiotherapy quality.
AB - Introduction Modern technologies, like intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT), have improved the therapeutic ratio of thoracic radiotherapy (TRT) for lung cancer (LC). Halcyon™ (Varian Medical Systems, Palo Alto, CA, USA), a novel 6MV-flattening-filter-free O-ring linear accelerator (6X-FFF ORL), was designed to deliver IMRT and VMAT with greater speed than a C-arm linac. Herein, we report our initial clinical experience treating patients with LC on this linac. Methods All patients who received TRT for LC on the 6X-FFF ORL at our institution were retrospectively identified. Patients' clinicopathologic data, radiotherapy details, early disease-control and toxicity outcomes, dosimetric data, couch corrections, and treatment times are reported. Results Between 10/2018-12/2019, 30 consecutive patients (median age 66 years, range 54-94 years) received definitive or post-operative TRT for LC (median 66 Gy/33 fractions; range 5-70 Gy/2-37 fractions) following four-dimensional computed tomography (CT) simulation (97%) using daily kilovoltage KV cone-beam CT (CBCT) (100%) on a 6X-FFF ORL for non-small cell LC (84%) or small cell LC (16%), with 53% receiving VMAT, 43% receiving static-field IMRT, and 77% receiving concurrent systemic therapy. All plans were approved through institutional peer review. The average three-dimensional vector couch correction based on CBCT guidance was 0.90 ± 0.50 cm. The average beam-on and beam on plus CBCT times were 1.7 ± 1.1 min, and 5.0 ± 3.2 min, respectively. Grade 3 dyspnea and fatigue occurred in 3% and 3% of patients, respectively. There were no grade ≥4 toxicities. Conclusion In this first clinical report of TRT for LC on a 6X-FFF ORL, daily CBCT-guided treatment was fast and safe with respect to dosimetry and clinical outcomes. Thus, use of this linac for TRT may increase LC patient throughput without a detriment in radiotherapy quality.
KW - Flattening filter free
KW - Imrt
KW - Lung cancer
KW - Non-small cell lung cancer
KW - Radiation
KW - Radiotherapy
KW - Small cell lung cancer
KW - Throughput
KW - Vmat
UR - https://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=purepublist2023&SrcAuth=WosAPI&KeyUT=WOS:000567293300004&DestLinkType=FullRecord&DestApp=WOS
U2 - 10.7759/cureus.10325
DO - 10.7759/cureus.10325
M3 - Paper
C2 - 33052286
SP - e10325