Skip to main navigation Skip to search Skip to main content

High intensity focused ultrasound for the treatment of rat tumours

  • G ter Haar
  • , I Rivens
  • , L Chen
  • , S Riddler
  • Royal Marsden Hospital

Research output: Contribution to journal › Article › peer-review

105 Scopus citations

Abstract

Discrete implanted liver tumours in the rat have been exposed to arrays of 1.7 MHz ultrasound lesions. Focal peak intensities in the range 1.4-3.5 k Wcm-2 were used for an exposure time of 10 s. It has been demonstrated that where the whole tumour volume was exposed to the focused ultrasound beam, no evidence of tumour growth could be detected histologically. Where the ultrasonic lesion array was not contiguous, regrowth occurred. Preliminary histological studies confirmed this finding.

Original languageEnglish
Article number009
Pages (from-to)1495-1501
Number of pages7
JournalPhysics in Medicine and Biology
Volume36
Issue number11
DOIs
StatePublished - Nov 1991

Keywords

  • Animals
  • Female
  • Fibrosarcoma/therapy
  • Neoplasm Transplantation
  • Rats
  • Rats, Inbred Strains
  • Ultrasonic Therapy

Fingerprint

Dive into the research topics of 'High intensity focused ultrasound for the treatment of rat tumours'. Together they form a unique fingerprint.

Cite this