Correction factors for Fricke dosimetry in high-energy electron beams

C. M. Ma, A. E. Nahum

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

The paper presents Fricke-to-water dose conversion and wall correction factors for Fricke ferrous sulphate dosimetry in high-energy electron beams. The dose conversion factor has been calculated as the ratio of the mean dose in water to the mean dose in the Fricke solution with a water-walled vessel and the wall correction factor accounts for the change in the Fricke dose due to the presence of the non-water wall material. The EGS4 (electron gamma shower version 4) Monte Carlo code system has been employed in the work together with the application of a correlated sampling variance reduction technique. The results show that for a Fricke dosimeter of 1.534 cm diameter and 5.5 cm length the dose conversion factor is nearly constant at 1.004 (within 0.1%) for electron energies of 11-25 MeV if the dosimeter is placed at the depth of maximum dose but can vary by a few per cent if the dosimeter is placed on the descending portion of the depth-dose curve.

Original languageEnglish
Article number009
Pages (from-to)423-438
Number of pages16
JournalPhysics in Medicine and Biology
Volume38
Issue number3
DOIs
StatePublished - 1993

Fingerprint

Dive into the research topics of 'Correction factors for Fricke dosimetry in high-energy electron beams'. Together they form a unique fingerprint.

Cite this