A rapid method to regenerate piezoelectric microcantilever sensors (PEMS)

Lina Loo, Wei Wu, Wan Y. Shih, Wei Heng Shih, Hossein Borghaei, Kambiz Pourrezaei, Gregory P. Adams

Research output: Contribution to journalArticlepeer-review

11 Scopus citations

Abstract

Piezoelectric microcantilever sensors (PEMS) can be sensitive tools for the detection of proteins and cells in biological fluids. However, currently available PEMS can only be used a single time or must be completely stripped and refunctionalized prior to subsequent uses. Here we report the successful use of an alternative regeneration protocol employing high salt concentrations to remove the target, leaving the functional probe immobilized on the microcantilever surface. Our model system employed the extracellular domain (ECD) of recombinant human Epidermal Growth Factor Receptor (EGFR) as the probe and anti-human EGFR polyclonal antibodies as the target. We report that high concentrations of MgCl2 dissociated polyclonal antibodies specifically bound to EGFR ECD immobilized on the sensor surface without affecting its bioactivity. This simple regeneration protocol both minimized the time required to re-conjugate the probe and preserved the density of probe immobilized on PEMS surface, yielding identical biosensor sensitivity over a series of assays.

Original languageEnglish
Pages (from-to)5520-5528
Number of pages9
JournalSensors
Volume11
Issue number5
DOIs
StatePublished - May 2011

Keywords

  • Antibody
  • Antigen
  • Biomarker
  • Biosensor
  • Piezoelectric microcantilever
  • Regeneration

Fingerprint

Dive into the research topics of 'A rapid method to regenerate piezoelectric microcantilever sensors (PEMS)'. Together they form a unique fingerprint.

Cite this